| SPECIFICATIONS |
| Specifications
apply after 30 minute warm-up, ambient 5?C to 40?C |
| FREQUENCY |
| Frequency Range: |
10MHz to 1000MHz |
| Setting
Resolution: |
1kHz by direct keyboard entry,
or in user-set increments of 1kHz to 999?999MHz by rotary control or
increment?decrement keys. |
| Display
Resolution: |
1kHz |
| Accuracy: |
? 2 ppm over temperature
range 5?C to 40?C. |
| Stability: |
<1ppm/year ageing. |
| Phase Noise: |
-110dBc/Hz at 25kHz offset,
500MHz carrier. |
| Residual FM (FM
Off): |
: |
| Equivalent peak
deviation for 300Hz to 3?4kHz B/W: |
10Hz at 100MHz carrier, 35Hz
at 500MHz carrier, 180Hz at 1000MHz carrier |
| OUTPUT
LEVEL |
| Output Level
Range: |
?127dBm to +7dBm (0?1?V to
500mV into 50W). |
| Setting
Resolution: |
0?1dB (or 0?01?V to 1mV) by
direct keyboard entry, or in user-set increments of 0?1dB to 100dB (or
0?01?V to 100mV) by rotary control or increment?decrement keys. |
| Accuracy: |
Better than ? 2dBm, except
for output levels <-70dBm at 500 -1000 MHz, ? 3dBm. |
| Harmonics: |
Typically <-25dBc, maximum
?20dBc, any carrier frequency, output level <0dBm. |
| Non-Harmonic
Spurii: |
<?60dBc at >8kHz
offset. |
| Carrier Leakage: |
<0.5uV generated into a 50W
load by a 2 turn 25mm loop, at 25mm from the generator with output set
to <-10dBm into a 50W
sealed load. |
| Output Impedance: |
50W |
| Output Connector: |
TYPE N |
| Reverse
Protection: |
50V DC |
| Output Switch: |
RF OUT on-off switch with LED
for ON status. |
| FM
MODULATION |
| Peak Deviation: |
0?5kHz to 100kHz. |
| Setting
Resolultion: |
0?5kHz by direct keyboard
entry, rotary control or increment?decrement keys. |
| Modulation
Frequency: |
Internal 1kHz; External 300Hz
to 50kHz |
| Deviation
Accuracy: |
<?10% of setting ?0.5kHz,
excl. residual FM, for 1kHz modulation, internal or 1Vrms external. |
| External
Modulation |
|
| Frequency
Response: |
? 1dB from 30Hz to 50kHz
relative to 1kHz. |
| Distortion: |
<2% total harmonic
distortion at 1kHz mod. frequency, 100kHz deviation and 500MHz carrier. |
| Input Impedance: |
100kOhm |
| Input Connector: |
BNC |
| INTERFACES |
| Full
remote control facilities are available through the RS232 (standard) or
optional GPIB interfaces. |
| RS232: |
Variable Baud rate, 19200 Baud
maximum, 9-pin D-connector. Compatible with Thurlby Thandar ARC
(Addressable RS232 Chain). |
| IEEE-488: |
Conforming with IEEE488?1 and
IEEE488?2. |
| GENERAL |
| Display: |
20 character x 4 row backlit
alphanumeric LCD |
| Data Entry: |
Keyboard selection of
frequency, amplitude, etc.; value entry by numeric keys or by rotary
control. |
| Stored Settings: |
Up to 9 complete instrument
set-ups may be stored in battery-backed memory. |
| Size: |
3U (130mm) height; half-rack
(212mm) width; 330mm long. |
| Weight: |
4?6 kg. (10 lb) |
| Power: |
230V, 115V or 100V nominal
50/60Hz, adjustable internally; operating range ?14% of nominal; 30VA
max. Installation Category II. |
| Options: |
IEEE-488 interface; 19 inch
rack mounting kit. |
| ENVIRONMENTAL
& SAFETY |
| Operating Range: |
+5?C to 40?C, 20-80% RH. |
| Storage Range: |
?20?C to + 60?C. |
| Environmental: |
Indoor use < 2000m,
Pollution Degree 1. |
| Safety: |
Complies with EN61010-1. |
| EMC: |
Complies with EN61326 |