  1906  - Description and Specifications

1906 - 5.5 Digit LED Bench Multimeter

  5˝ digits (±210,000 counts), auto/manual range
  1µV, 1nA, 1mOhm resolution, 0.012% accuracy
  Wide range of computing functions included
  Automatic data logging functions included
  Integral True RMS AC converter
  Isolated addressable RS232 interface standard
  Model with GPIB (IEEE-488) interface available
 

  Larger Image  

Greater resolution and accuracy
The 1906 is a true 5.5 digit meter with a scale length of ± 210,000 counts. This gives it not just superior resolution relative to 4.5 digit meters, but much higher effective accuracy as well.
Greater sensitivity and higher input impedance
With a maximum resolution of 1mV, 1mW and 1nA, the 1906 is ten times more sensitive than a 4.5 digit meter.
This extra senstivity enables accurate measurements to be made in areas previously impossible such as thermocouple junctions, switch contact resistance or capacitor leakages.
4 terminal Ohms for improved stability
Resistance measurements can be made in either 2 terminal or 4 terminal mode. High impedance sensing ensures that the meter switches automatically between modes. For low resistance measurements, 4 terminal mode offers much higher accuracy and repeatability.
True RMS ac functions
The 1906 provides True RMS ac response which gives accurate measurements regardless of the waveform shape. The wide bandwidth voltage attenuator provides high accuracy within the audio band and gives extended response to avoid distortion when measuring switching waveforms.
Full bus control via RS-232 or GPIB
The 1906 is designed for complete bus control. Every function of the instrument can be controlled using either the RS-232 interface or the GPIB interface (optional).
Wide range of Smart functions
The 1906 offers computing and datalogging functions as standard: dBs, Ax+b, Limits, % deviation, Min-Max, and a 100 reading Data-logger.
Auto or manual ranging
The 1906 offers fully automatic ranging on all functions including current. Alternatively any range can be selected and held manually.
Closed case calibration
The 1906 stores its calibration constants in permanent memory. Consequently it can be recalibrated without ever opening the case.
Storage of front-panel set-ups
The 1906 can store up to six front panel set-ups in non-volatile memory. In addition to the range and function each set-up defines the active computing or logging functions along with their parameters.
Single key convenience functions for ease of use
Press the “null” key and the present displayed reading is stored and subtracted from all future readings. This powerful feature has a host of uses such as removing test lead resistance, observing deviation levels, or measuring relative to alternative voltage points using a fixed ground lead connection.
Full protection for the instrument and the user
The 1906 incorporates extensive protection against damage from accidental overloads on all ranges including 10 Amps.
In addition the 1906 meets the stringent safety requirements of IEC348 and IEC1010-1 for measurements up to 1kV.


Touch and Hold
The touch and hold function is invaluable when taking measurements in awkward situations where it may be difficult to see the multimeter display. With touch and hold selected, the display is automatically frozen every time a stable reading is achieved. A further press of the T/HOLD key un-freezes the display.
Digital filtering for better noise performance
The 1906 uses non-linear digital filtering which is highly effective in eliminating unwanted noise, but responds rapidly to a change in the reading. When working with very noisy or unstable measurements, the characteristics of the digital filter can be adjusted to suit the application.
Linear scaling with offset (Ax+b)
This function enables a reading to be multiplied by a scale factor (A) and for an offset (b) to be added or subtracted. This means, for example, that the electrical output of a transducer can be scaled to give a direct readout of the physical parameter e.g. temperature or weight. It is particularly useful with 4-20mA current loops.
Limits comparison
This function enables the reading to be tested against high and low limits set by the user. The display then shows a code of H (high), L (low) or P (pass) at the end of the display.
Percentage deviation (D%)
This function enables the meter to display the percentage amount by which the reading differs from a nominal value chosen by the user. It is invaluable for measuring tolerance or stability.
Min-Max storage
The Min-Max function stores the highest and lowest values of a set of readings. This facility has many uses such as checking for power supply glitches, recording peak temperature excursions and avoiding “missed” readings when using the Logger function.
Logarithmic measurements (dB)
This function allows measurements to be displayed in the form 20log10X. Built-in zero reference scaling enables voltages to be displayed in dBV or dBm relative to any required impedance. Using the null key allows gains and losses to be measured directly.
Automatic data logging
The 1906 can store up to 100 readings at any required time interval from 1 reading per second up to 1 reading every 3 hours.
A simple recall sequence allows the readings to be scrolled onto the display whenever required. Alternatively results can be downloaded using the RS232 or GPIB interfaces. As a result, tedious time related measurement sequences can be handled automatically without disrupting your work.
Manual storage of readings is also available eliminating the need for paper and pencil when making a series of measurements. Storage can also be triggered from the RS-232 or GPIB interfaces.

ACCURACY
Accuracies apply for 1 year 18oC to 26oC. Temp. coefficient outside these limits is <0×1 x quoted range
accuracy per oC. Warm-up time to rated accuracy is 1 hour.

DC VOLTS

Range

Accuracy(±)

Input Impedance

Max Input

200mV

0×017% + 3d

10M or >1G (Hi Z mode)

1kV DC or AC Peak any range   

2V

0×012% + 3d

 

20V

0×019% + 3d

10M

200V

0×019% + 3d

 

1000V

0×019% + 3d

 

RESISTANCE (OHMS)

Range

Accuracy(±)

Max. Measuring Current

Max. Input

200

0×025% + 4d

1×5mA

300V DC or RMS continuous any Ohms input any range

2k

0×019% + 3d

1×5mA

20k

0×019% + 3d

0×4mA

200k

0×019% + 3d

40mA

2M

0×022% + 3d

4mA

20M

0×07% + 3d

400nA

Max open circuit voltage 3×5V.  Diode Test  on 2k range.
Audible continuity check available on all ranges.

AC VOLTS (True RMS)

Range

Accuracy(±)

 

45Hz - 5kHz

5kHz - 10kHz

10kHz-20kHz

20kHz-50kHz

200mV

0×2% + 150d

0×2% + 250d

3% + 500d

-

2V

0×2% + 150d

0×2% + 250d

0×5% + 300d

3% + 1500d

20V

0×2% + 150d

0×2% + 250d

0×5% + 300d

2% + 1000d

200V

0×2% + 150d

0×2% + 250d

0×5% + 500d

-

750V

0×2% + 150d

-

-

-

Additional error at crest factor=3 typically 0×7%, Input impedance 1M / <100 pF any range.
Max input 750V rms, 1kV peak, any range.

DC AMPS

Range

 

Accuracy(±)

Voltage Burden

Max. Input

200mA

 

0×08% + 12d

300 mV max

1A, 300V fuse protected

2mA

 

0×08% + 12d

 

20mA

 

0×08% + 12d

 

200mA

 

0×08% + 12d

 

10A

 <2000mA

0×08% + 12d

650mV max

10A, 300V fuse protected

 

>2000mA

0×12% + 12d

 

AC AMPS (True RMS)

Range

Accuracy(±) 45 Hz - 1 kHz

Voltage Burden

Max Input

200mA

 

0×37% + 100d

300mV max

1A, 300V fuse protected

2mA

 

0×37% + 100d

 

20mA

 

0×37% + 100d

 

200mA

 

0×37% + 100d

 

10A

<2000mA

0×37% + 100d

650mV max

10A, 300V fuse protected

 

>2000mA

0×4% + 100d

 

Additional error at crest factor=3 typically 0×7%.

FURTHER SPECIFICATIONS
INPUT CHARACTERISTICS
Input Current: < 100 pA.
DC NMR: > 60dB at 50/60Hz.
1k Unbalanced CMR: > 120dB at DC/50Hz/60Hz, DC ranges; > 60dB at DC/50Hz/60Hz, AC ranges.
DISPLAY
Display Type: 13mm LED, 8 digit.
Scale Length: Selectable 5˝ digit or 4˝ digit.
Reading Rate: 3 readings/sec (5˝ digit). 10 readings/sec (4˝ digit).
Overrange: Displays OL if input too great for range.
Overflow: Displays OFLO if calculated result overflows display.
Annunciators: LEDs for range, function and program modes.
COMPUTING FUNCTIONS
Null: Operates over full range; values can be stored for every function.
Digital Filter: 10 options, including optimised default values for each range and function.
Touch & Hold: Reading is held when stable.
% DEV: Displays % deviation from entered reference value.
dB: Displays measurement in dB relative to 1V, 1mA, 1kW or user entered value, or in dBm.
Ax+B: Linear scaling of results, with offset.
Limits: Reading displayed with H, L, or P (pass) with respect to user-defined high and low limits.
Min/Max: Minimum and maximum reading stored.
Data Logger: Manual or automatic storage of 100 measurements, Storage interval 1s to 9999s.
INTERFACES
RS232: Baud rates 300, 1200 or 9600. Complies fully with the ARC (Addressable RS232 Chain) interface standard. Address selectable from the front panel.
GPIB (Optional): Fully complies with IEEE-488.1 and IEEE-488.2
POWER REQUIREMENTS
AC Input: 110/120 Volts AC nominal 50/60Hz or 220/240 Volts AC nominal 50/60Hz by rear panel adjustment. The instrument will operate safely and meet specification within normal AC supply variations, viz. 99-132 Volts AC and 198-264 Volts AC respectively.
Consumption: 25VA max.
GENERAL
Operating Range: +5oC to +40oC, 20% to 80% RH
Storage Range: -40oC to +60oC
Size: 260(W) x 88(H) x 235(D)mm, excl. handle and feet
Weight: 2×2kg.