All
accuracies apply for 1 year, 19oC to 25oC. Temperature
coefficient outside these limits is <0.1 x quoted range accuracy per oC.
DC
VOLTAGE
|
Range
|
Accuracy
|
Resolution
|
Notes
|
|
400mV
|
0.08%
± 4 dig.
|
10µV
|
Max.
Input 265V DC/AC rms
|
|
4V
|
0.08%
± 4 dig.
|
100µV
|
Input
impedance 10MW nominal Max. input 1kV DC or AC pkNMR: >60dB @
50/60HzCMR: >90dB @ DC/50Hz/60Hz
|
|
40V
|
0.08%
± 4 dig.
|
1mV
|
|
400V
|
0.08%
± 4 dig.
|
10mV
|
|
1000V
|
0.09%
4 dig.
|
100mV
|
AC
VOLTAGE (True
RMS, 4000 count scale length)
|
Range
|
Accuracy
|
Resolution
|
|
|
45Hz
- 400Hz
|
400Hz
- 5kHz
|
5kHz
- 20kHz
|
|
|
400mV
|
0.5%
± 4 dig.
|
1%
± 4 dig.
|
3%
± 4 dig.
|
100µV
|
|
4V
|
2%
± 4 dig.
|
5%
± 4 dig.
|
1mV
|
|
40V
|
10mV
|
|
400V
|
3%
± 4 dig.
|
100mV
|
|
750V
|
1%
± 4 dig.
|
|
1V
|
Accuracies
apply for readings between 10% and 100% of full scale. Additional error
at crest factor = 3 is typically 1%.
Input impedance = 10MOhms nominal. Max. input = 750V rms, 1kV
pk. (265V rms on 400mV range). 1kOhm
unbalanced CMR = >60dB at DC or 50Hz (60Hz rejection available as
factory option).
RESISTANCE
|
Range
|
Accuracy
|
Resolution
|
Notes
|
|
400W
|
0.15%
± 6 dig.*
|
10mW
|
*
400Ohm specification applies
after null Max. input 265V DC or AC rms on any range
.Max. open
circuit voltage 4V
40MOhm accuracy
applies up to 20M thereafter add 1%
|
|
4kW
|
0.1%
± 4 dig.
|
100mW
|
|
40kW
|
0.1%
± 4 dig.
|
1W
|
|
400kW
|
0.15%
± 4 dig.
|
10W
|
|
4MW
|
0.3%
± 6 dig.
|
100W
|
|
40MW
|
1.0%
± 10 dig.
|
1kW
|
DC
CURRENT
|
Range
|
Accuracy
|
Resolution
|
Notes
|
|
4mA
|
0.1%
± 4 dig.
|
0.1µA
|
Max.
input 1A (Fused)
Voltage burden <500mV
|
|
400mA
|
0.1%
± 4 dig.
|
10µA
|
|
10A
(up to 1A)
|
0.3%
± 4 dig.
|
1mA
|
Max. input 10A (Fused)
Voltage burden <500mV
|
|
10A
(up to 5A)
|
1.0%
± 4 dig.
|
1mA
|
|
10A
(up to 10A)
|
3.0%
± 10 dig.
|
1mA
|
AC
CURRENT (True
RMS, 4000 count scale length)
|
Range
|
Accuracy
|
Resolution
|
Notes
|
|
1mA
|
0.5%
± 4 dig.
|
1µA
|
Max.
input 1A (Fused)
Voltage burden <500mV
|
|
100mA
|
0.5%
± 4 dig.
|
100µA
|
|
10A
(up to 1A)
|
0.8%
± 4 dig.
|
10mA
|
Max. input 10A (Fused)
Voltage burden <500mV
|
|
10A
(up to 5A)
|
1.5%
± 4 dig.
|
10mA
|
|
10A
(up to 10A)
|
3.0%
± 4 dig.
|
10mA
|
Accuracies
apply over 45Hz to 10kHz for readings between 10% and 100% of range.
Additional error at crest factor = 3 is typically 1%.
FURTHER
FUNCTIONS
Continuity:
Selects 4kW
range and sounds audible tone for impedance <10W.
Max. input 265V DC or AC rms.
Diode
Test: Displays junction voltages up to 3V at a test current of 1mA
at 1V. Max. open
circuit voltage approximately 4V. Max. input 265V DC or AC rms.
Null:
(Relative): Stores
current reading and subtracts it from future readings.
Hold:
Reading is
frozen until released.
T-Hold:
Reading is
frozen when it becomes stable.
Min/Max:
Minimum and
maximum readings are stored.