FREQUENCY
|
Frequency Range:
|
0.005Hz to 5MHz in 7 overlapping decade
ranges with fine adjustment by verniers.
|
Vernier Range:
|
1000:1 on each range.
|
FREQUENCY LOCKING
|
Operating Range:
|
0.5Hz to 5MHz with vernier setting within
10% to 100% of range maximum.
|
Locking Accuracy:
|
Better than 0.01% of displayed value.
|
DISPLAY
|
LCD, 8 digits (11mm high) plus 14
annunciators
|
METER FUNCTIONS (generator
mode)
|
Frequency:
|
Auto-ranging reciprocal measurement giving 4
digit resolution for frequencies down to 1Hz; maximum resolution is
0.001Hz. Accuracy ±1 digit 0.2Hz to 5MHz; accuracy ±1% of range below
0.2Hz
|
Amplitude:
|
Display shows peak-to-peak amplitude or rms
value. Display corrected for attenuator setting. 3-digit resolution,
accuracy typically 5% of range
|
DC Offset:
|
3-digit resolution; accuracy typically ±2%
of setting ±1 digit. Display corrected for attenuator setting
|
OUTPUTS
|
MAIN - 50 Ohm
|
Amplitide
|
2mV to 20V peak-peak open
circuit (1mV to 10V peak-peak into 50Ohms) in four switch-selectable ranges
with 20dB vernier control within each range
|
Attenuator Ranges:
|
0dB, -20dB, -40dB, -60dB
|
DC Offset Range:
|
±10V from
50W. DC offset plus signal peak
limited to ±10V (±5V into 50W). DC offset plus waveform attenuated
proportionally by the attenuator.
|
MAIN - 600 Ohm
|
Alternative output socket
offering the same facilities as the 50W socket.
|
AUX OUT
|
0 to 5V TTL/CMOS logic levels
capable of driving 2 standard TTL loads. Frequency, symmetry and phase as
main outputs
|
SWEEP OUT
|
3V ramp from
600W
|
EXTERNAL COUNTER
|
Frequency Range:
|
5Hz to 20MHz, fully autoranging
|
Input Sensitivity:
|
50mV rms (sinewave)
|
Measurement Time:
|
Selectable 0.5s or 5s
|
Resolution:
|
6 digits in 0.5s; 7 digits in 5s
|
Accuracy:
|
±1 digit ± timebase accuracy
|
Timebase Accuracy:
|
±10ppm initial error; ±5ppm/year ageing
rate; typically less then 0.5ppm/oC. Adjustment point for closed-case
recalibration
|
OPERATING MODES
|
Specifications apply for the top
decade of each frequency range and maximum output into 50W termination.
|
SINE
|
Distortion:
|
<0.5% on 500, 5k and 50k ranges; <1%
on 5, 50 and 500k ranges; all harmonics >25dB below fundamental on 5M
range.
|
Amplitude Flatness:
|
±0.2dB to 200kHz; ±2dB to 5MHz.
|
TRIANGLE
|
Linearity:
|
Better than 99% to 200kHz
|
SQUARE WAVE
|
Rise and Fall Times:
|
<45ns
|
Mark - Space Ratio:
|
1:1 ± 1% to 100kHz
|
DC
|
Range:
|
±10V unterminated
|
SYMMETRY
|
Symmetry Range:
|
Variable typically 1:9 to 9:1 (on top decade
of each range), frequency divided by 10.
|
SWEEP MODES
|
INTERNAL
|
Sweep Range:
|
>1000:1 within each range.
|
Sweep Rate:
|
Adjustable, typically 20ms to 20 secs.
|
Sweep Mode:
|
Linear or logarithmic. Sweep start and stop
frequencies displayed by button press.
|
EXTERNAL
|
Input Impedance:
|
10kW
|
Input Sensitivity:
|
0 to 3V for 1000:1 sweep
|
Max. Input Voltage:
|
±10V
|
Sweep Linearity:
|
Better than 1%
|
Max. Voltage Rate:
|
0.1V/us
|
AMPLITUDE MODULATION
|
Depth:
|
Variable 0 to 100%
|
Frequency:
|
400Hz (internal). DC to 100kHz (external).
|
External Sensitivity:
|
Approx. 2V peak-to-peak for 50% modulation.
|
GENERAL
|
POWER REQUIREMENTS
|
Input Voltage:
|
230V or 115V nominal 50/60Hz, by internal
adjustment; operating range ±14% of nominal. Installation Category II.
|
Consumption:
|
25VA max.
|
ENVIRONMENTAL & MECHANICAL
|
Operating Range:
|
+5oC to
+40oC, 20% to 80% RH.
|
Storage Range:
|
-10oC to
+65oC
|
Environmental:
|
Indoor use at <2000m, Pollution degree I
|
Safety:
|
Complies with EN601010-1
|
EMC:
|
Complies with EN50081-1 & EN50082-1
|
Size:
|
260(W) x 88(H) x 235(D)mm (10.2 x 3.4 x
9.2"), excluding handle and feet.
|
Weight:
|
1.9kg (4.2lb) approx.
|